Table 9-7: Adult Ventilator Alarm Settings / Status<br />

Mode: Upon Request<br />

Length: 14 Bytes<br />

Format: [ SYN ] [ID] [Data #1] [Data #13] [Check]<br />

[ ID ] : 40h<br />

[ Data #1 ] Alarm Status Byte - 1<br />

Bit 7 — High Min Vol (1=on, 0=off)<br />

Bit 6 — Low Min Vol (1=on, 0=off)<br />

Bit 5 — High Insp Pres (1=on, 0=off)<br />

Bit 4 — Low Insp Pres (1=on, 0=off)<br />

Bit 3 — High Base Pres (1=on, 0=off)<br />

Bit 2 — Low Base Pres (1=on, 0=off)<br />

Bit 1 — High Rate (1=on, 0=off)<br />

Bit 0 — Low Rate (1=on, 0=off)<br />

[ Data #2 ] Alarm Status Byte - 2<br />

Bit 7 — I:E Limit (1=on, 0=off)<br />

Bit 6 — Run Diagnostics (1=on, 0=off)<br />

Bit 5 — Gas Fail (1=on, 0=off)<br />

Bit 4 — Failed to Cycle (1=on, 0=off)<br />

Bit 3 — Reserved (0)<br />

Bit 2 — Reserved (0)<br />

Bit 1 — Reserved (0)<br />

Bit 0 — Alarm Silence (0)<br />

[Data #n] Parameter Range Binary Scale Binary Range<br />

[Data #3] High Min Vol 0 - 80 L 1 L 0 - 80<br />

[Data #4] Low Min Vol 0 - 50 L 1 L 0 - 50<br />

[Data #5] High Insp Pres 0 - 120 cmH2O 1 cmH2O 0 - 120<br />

[Data #6] Low Insp Pres 3 - 99 cmH2O 1 cmH2O 3 - 99<br />

[Data #7] High Base Pres 0 - 55 cmH2O 1 cmH2O 0 - 55<br />

[Data #8] Low Base Pres 0 - 50 cmH2O 1 cmH2O 0 - 50<br />

[Data #9] High Rate 0 - 155 bpm 1 bpm 0 - 155<br />

[Data #10] Low Rate 3 - 99 bpm 1 bpm 3 - 99<br />

[Data #11] I:E Override on/off n/a 1/0<br />