Service <strong>Manual</strong> 35<br />

OVP Checklist<br />

This checklist is for use during the BEAR CUB 750vs Operational Verification Procedure.<br />

Serial Number Hour meter reading Today’s date<br />

Service Location Service Organization<br />

Address Address<br />

City/State/Zip City/State/Zip<br />

Contact Contact<br />

Phone ( ) Phone ( )<br />

Verification Steps<br />

STEP PASS FAIL STEP PASS FAIL<br />

Indicator Verification O2 Pressure Reading<br />

Verification<br />

Breath Rate Verification Low Gas Supply<br />

Verification(Air)<br />

Alarm Silence, Reset Verification Low Gas Supply Verification<br />

(O2)<br />

Exhalation Time, I:E Ratio<br />

Verification<br />

Fail to Cycle alarm<br />

verification<br />

Low Pressure alarm Verification FIO2 verification (21%, 60%,<br />

100%)<br />

High Pressure alarm/Limit<br />

Verification<br />

Low PEEP/CPAP alarm<br />

verification<br />

Pressure LED reset AC power loss verification<br />

Prolonged Inspiratory alarm<br />

verification<br />

Tidal Volume Limit<br />

verification<br />

High Rate alarm verification Flow Sensor alarm<br />

verification<br />

Apnea alarm verification Patient Triggering verification<br />

Incompatible setting alarm<br />

verification<br />

Air Pressure reading verification<br />

ADD Warning ?????<br />

Signature _________________________________________<br />

PROCEDURE COMPLETE<br />

Alarm loudness verification<br />

L2342 Revision B November 2004