Service <strong>Manual</strong> 39<br />
Table 4-2: Calibration<br />
PATIENT port. To calibrate the Flow Control Valve, press the TEST key to<br />
advance the calibration number; then set flow as specified and press the<br />
MANUAL BREATH key to record data. Calibrate Base Flow, then Calibrate<br />
Inspiratory Flow.<br />
Transducer Calibration Number Message Input <strong>Static</strong> Pressure<br />
Differential (port B) d10 0 ± 1 0 cmH2O<br />
d10 50 ± 1 50 ± 0.5 cmH2O<br />
Machine (port B) d11 0 ± 1 0 cmH2O<br />
d11 50 ± 1 50 ± 0.5 cmH2O<br />
Proximal (port B) d12 0 ± 1 0 cmH2O<br />
d12 50 ± 1 50 ± 0.5 cmH2O<br />
AIR (A1) d13 50 ± 2 50 ± 0.2 PSI<br />
O2 (A2) d14 50 ± 2 50 ± 0.2 PSI<br />
NOTE<br />
The pressure display occasionally flickering 1 digit beyond the tolerance is<br />
acceptable.<br />
L2342 Revision B November 2004