Service <strong>Manual</strong> 39<br />

Table 4-2: Calibration<br />

PATIENT port. To calibrate the Flow Control Valve, press the TEST key to<br />

advance the calibration number; then set flow as specified and press the<br />

MANUAL BREATH key to record data. Calibrate Base Flow, then Calibrate<br />

Inspiratory Flow.<br />

Transducer Calibration Number Message Input <strong>Static</strong> Pressure<br />

Differential (port B) d10 0 ± 1 0 cmH2O<br />

d10 50 ± 1 50 ± 0.5 cmH2O<br />

Machine (port B) d11 0 ± 1 0 cmH2O<br />

d11 50 ± 1 50 ± 0.5 cmH2O<br />

Proximal (port B) d12 0 ± 1 0 cmH2O<br />

d12 50 ± 1 50 ± 0.5 cmH2O<br />

AIR (A1) d13 50 ± 2 50 ± 0.2 PSI<br />

O2 (A2) d14 50 ± 2 50 ± 0.2 PSI<br />

NOTE<br />

The pressure display occasionally flickering 1 digit beyond the tolerance is<br />

acceptable.<br />

L2342 Revision B November 2004