Service <strong>Manual</strong> 43<br />

Table 4-6: Standard Test Settings<br />

CONTROLS<br />

Mode Assist/Control<br />

PEEP/CPAP Pressure 0 cmH2O<br />

Inspiratory Pressure (PIP) 20 cmH2O<br />

Ventilator Rate 30 BPM<br />

Inspiratory Time 1.00 second<br />

Volume Limit 300 mL<br />

Base Flow 5 L/min<br />

Inspiratory Flow 15 L/min<br />

Assist Sensitivity Mid Range<br />

ALARMS<br />

Low PEEP/CPAP Alarm -5 cmH2O<br />

High Breath Rate Alarm 255 BPM<br />

Low Inspiratory Pressure Alarm 5 cmH2O<br />

High Pressure Limit Alarm 75 cmH2O<br />

FRONT OF VENTILATOR<br />

Humidifier VH-820 chamber, no water<br />

Patient Circuit Standard Allied Infant Circuit<br />

Patient Resistance Rp 20<br />

Patient Compliance C3<br />

Endotracheal Tube None<br />

Auxiliary Pressure Relief Fully closed (CW)<br />

REAR OF VENTILATOR<br />

Alarm Loudness Control Minimum<br />

RS-232 Interface Open<br />

Analog Output Open<br />

Remote Nurses call Open<br />

APNEA Alarm 30 seconds<br />

Display Power Outlet Open<br />

AC Line Voltage 120 VAC<br />

O2 Supply >35 PSI<br />

Air Supply >35 PSI<br />

L2342 Revision B November 2004