98 Bear Cub 750PSV<br />
Table 9-6: Adult Ventilator Control Settings<br />
Mode: Upon Request<br />
Length: 16 Bytes<br />
Format:<br />
[ SYN ] [ID] [Data #1] [Data #13] [Check]<br />
[ ID ] : 20h<br />
[ Data #1 ] Mode/Wave:<br />
Bit 7 — x<br />
Bit 6 — x<br />
Bit 5 — x<br />
Bit 4 — y<br />
Bit 3 — y<br />
Bit 2 — Reserved (0)<br />
Bit 1 — Reserved (0)<br />
Bit 0 — Reserved (0)<br />
Mode :<br />
000 — Assist Control<br />
001 — SIMV/CPAP/PSV<br />
010 — Pressure Control<br />
011 — Reserved<br />
: :<br />
111 — Reserved<br />
Wave:<br />
00 — Square<br />
01 — Sine<br />
10 — Decelerating<br />
11 — Reserved<br />
[Data #2] Switches:<br />
Bit 7 — 100% O2 (1=on, 0=off)<br />
Bit 6 — Sigh On (1=on, 0=off)<br />
Bit 5 — Pres Aug (1=on, 0=off)<br />
Bit 4 — Nebulizer (1=on, 0=off)<br />
Bit 3 — Reserved (0)<br />
Bit 2 — Reserved (0)<br />
Bit 1 — Reserved (0)<br />
Bit 0 — Reserved (0)<br />
[Data #n] Parameter Range Binary Scale Binary Range<br />
[Data #3] Tidal Vol 0.03 - 2.00L 0.01 L 3 - 200<br />
[Data #4] Rate 0-120 bpm 0.5 bpm 0 - 240<br />
[Data #5] Peak Flow 5 - 150 lpm 1 lpm 5 - 150<br />
[Data #6] Percent O2 21 - 100% O2 1% O2 21 - 100<br />
[Data #7] Assist Sens 0.2 - 5.0 cmH2O 0.1 cmH2O 2 - 50<br />
[Data #8] PSV/Insp Pres 0 - 80 cmH2O 1 cmH2O 0 - 80<br />
[Data #9] Insp Pause 0.0 - 2.0 sec 0.1 sec 0 - 20<br />
[Data #10] MMV 0 - 50 Liters 1 Liter 0 - 50<br />
[Data #11] Insp Time 0.0 - 5.0 sec 0.1 sec 0 - 50<br />
[Data #12] CC 0.0 - 7.5 ml/cmH2O 0.1 ml/cmH2O 0 - 75<br />
[Data #13] Pres Slope -P9 to P9, -9 to 9 1 1 – 38<br />
L2342 Revision B November 2004