Service <strong>Manual</strong> 19<br />

Breath Type (Patient Initiated)<br />

LED<br />

Test<br />

Push Button<br />

Battery<br />

LED On/Off<br />

Preset Values<br />

Minimum Expiratory Time 150 milliseconds<br />

Maximum Inspiratory Time 3.10 seconds<br />

Maximum settable I:E Ratio 4:1<br />

Model 9300<br />

Classification<br />

Degree of Protection<br />

Class 1<br />

Electric Shock Type B<br />

Harmful Ingress of Water<br />

Degree of Safety of Application<br />

None (Ordinary Equipment)<br />

in Presence of Flammable Anesthetics None<br />

Mode of Operation Continuous<br />

Emissions/Susceptibility<br />

This ventilator has been tested to conform to the following specifications:<br />

MIL-STD-461D:1993,MIL-STD-462D:1993, EN55011:1991, IEC 1000-4-2:1994, IEC<br />

1000-4-3:1994, IEC 1000-4-4:1994, IEC 1000-4-5:1994, QUASI-STATIC:1993<br />

This ventilator is designed and manufactured to comply with the safety requirements<br />

of IEC 601-1, IEC 601-2-12, CAN/CSA-C22.2 No. 601.1-M90, and UL 2601-1.<br />

Outputs<br />

Digital<br />

(see Section 10, Page 1) RS-232 Bi-Directional<br />

Analog<br />

Proximal Pressure -10 to 100 cmH2O, 1 cmH2O/25 mv<br />

L2342 Revision B November 2004