Service <strong>Manual</strong> 19<br />
Breath Type (Patient Initiated)<br />
LED<br />
Test<br />
Push Button<br />
Battery<br />
LED On/Off<br />
Preset Values<br />
Minimum Expiratory Time 150 milliseconds<br />
Maximum Inspiratory Time 3.10 seconds<br />
Maximum settable I:E Ratio 4:1<br />
Model 9300<br />
Classification<br />
Degree of Protection<br />
Class 1<br />
Electric Shock Type B<br />
Harmful Ingress of Water<br />
Degree of Safety of Application<br />
None (Ordinary Equipment)<br />
in Presence of Flammable Anesthetics None<br />
Mode of Operation Continuous<br />
Emissions/Susceptibility<br />
This ventilator has been tested to conform to the following specifications:<br />
MIL-STD-461D:1993,MIL-STD-462D:1993, EN55011:1991, IEC 1000-4-2:1994, IEC<br />
1000-4-3:1994, IEC 1000-4-4:1994, IEC 1000-4-5:1994, QUASI-STATIC:1993<br />
This ventilator is designed and manufactured to comply with the safety requirements<br />
of IEC 601-1, IEC 601-2-12, CAN/CSA-C22.2 No. 601.1-M90, and UL 2601-1.<br />
Outputs<br />
Digital<br />
(see Section 10, Page 1) RS-232 Bi-Directional<br />
Analog<br />
Proximal Pressure -10 to 100 cmH2O, 1 cmH2O/25 mv<br />
L2342 Revision B November 2004