Alarms<br />
Inspiratory Pressure<br />
Range: 0 to 72 cmH2O<br />
Resolution: 1 cmH2O from 0 to 30 cmH2O<br />
2 cmH2O from 30 to 72 cmH2O<br />
Repeatability: ± 1 cmH2O or ± 5% whichever is greater. Breath to breath<br />
<strong>Manual</strong> Breath<br />
Range: X1<br />
Assist Sensitivity<br />
Range: 0.2 to 5.0 L/min<br />
Resolution: 0.2 at 0.2 L/min increasing to 0.5 at 5.0 L/min<br />
Accuracy:Not a calibrated scale, minimum setting accuracy is ± 0.1 L/min<br />
Over Pressure Relief<br />
Range:<br />
to 75 cmH2O<br />
(Pressure at the patient wye): 15<br />
Resolution: 2 cmH2O stability (repeatability) ±4 cmH2O<br />
%02 Blending<br />
Range: 21 to 100% Oxygen concentration<br />
Resolution: 1%<br />
Accuracy: ± 3% oxygen<br />
APNEA Interval<br />
Range: 5, 10, 20, or 30 seconds<br />
Accuracy: ± 1 second<br />
Alarm Loudness<br />
Range: Min. to Max. 60 to 75 db(A)<br />
High Breath Rate<br />
Range: 3 to 255 BPM<br />
Low PEEP/CPAP<br />
Range: -5 to 30 cmH2O<br />
L2342 Revision B November 2004