Service <strong>Manual</strong> 93<br />

Chapter 9 RS-232 Protocol<br />

Scope<br />

This document defines both the hardware and software protocol for the RS-232<br />

communication port on the BEAR® 1000 Adult Ventilator, the BEAR CUB 750vs<br />

Infant Ventilator, and the BEAR® Graphics Display. Throughout this document the<br />

ventilator shall be defined as the SENDING device, that which transmits real time<br />

data and responds to requests for data in the form of other parameters. This is true<br />

even if the Graphic Display is in place acting as a repeater and providing data to a<br />

third device.<br />

Hardware Definition<br />

Connector: Male 9-pin DSub connector (on the Ventilator).<br />

Duplex: Full<br />

Baud Rate: Selectable 1200, 2400, 9600, or 19200<br />

Data: 8-bit character<br />

1 Start bit, 1 Stop bit<br />

No parity bit<br />

L2342 Revision B November 2004