Table 9-9: Infant Ventilator Control Settings<br />
Mode: Upon Request<br />
Length: 14 Bytes<br />
Format:<br />
[ SYN ] [ID] [Data #1] [Data #11] [Check]<br />
[ ID ] : A0h<br />
[ Data #1 ] Mode & Switches<br />
Bit 7 — x<br />
Bit 6 — x<br />
Bit 5 — x<br />
Bit 4 — Pressure Settings Incompatible (1=on, 0=off)<br />
Bit 3 — Flow Settings Incompatible (1=on, 0=off)<br />
Bit 2 — Rate/Time Settings Incompatible (1=on, 0=off)<br />
Bit 1 — <strong>Manual</strong> Breath (1=on, 0=off)<br />
Bit 0 — Reserved (0)<br />
Mode:<br />
000 — Assist Control<br />
001 — SIMV/IMV<br />
010 — CPAP<br />
011 thru 111 — Reserved<br />
[Data #n] Parameter Range Binary Scale Binary Range<br />
[Data #2] Breath Rate 0 - 150 BPM L 1 BPM 0 - 150<br />
[Data #3] Insp Time MSB 0.10 -3.00 sec 0.01 sec 10 - 300<br />
[Data #4] Insp Time LSB " " "<br />
[Data #5] Volume Limit MSB 5 - 300 mL 1 mL 5 - 300<br />
[Data #6] Volume Limit LSB " " "<br />
[Data #7] Insp Flow 1 - 30 L/min 0.5 L/min 2 - 60<br />
[Data #8] Base Flow 1 - 30 L/min 0.5 L/min 2 - 60<br />
[Data #9] Assist Sens 0.2 - 25.5 L/min 0.1 L/min 2 - 255<br />
[Data #10] Insp Pres Reserved 0<br />
[Data #11] PEEP/CPAP Reserved 0<br />