Service <strong>Manual</strong> 29<br />

Chapter 3 OVP Recommended Service<br />

Operational Verification<br />

Operational Verification should be performed between patient use or a minimum of<br />

once each month. Verification may be done by a qualified operator, Allied Healthcare<br />

Trained Hospital Service Technician or Allied Healthcare Technician. Verification is<br />

intended to be done in the hospital. If done by an Allied Healthcare Technician,<br />

there will be a charge.<br />

Certain procedures such as a verification of the alarm circuits should be performed at<br />

least once every 24 hours that the unit is in clinical use. Refer to the BEAR CUB<br />

750vs Infant Ventilator instruction manual for details.<br />

Do not use the ventilator unless it passes all the steps in the Operational Verification<br />

Procedure.<br />

Equipment Required<br />

Infant Test Lung P/N 52000-40027<br />

Patient Circuit P/N 50000-01147<br />

Oxygen Analyzer<br />

Rubber Stopper<br />

Stop Watch<br />

CAUTION<br />

If any of the following procedures fail to producer the results as outlined in this<br />

document, disconnect the ventilator and contact your authorized VIASYS Healthcare<br />

service technician.<br />

L2342 Revision B November 2004