Service <strong>Manual</strong> 41<br />

Table 4-4: Inspiratory Flow Calibration<br />

Flow Flow Flow Flow Flow Flow Flow Flow Flow<br />

Sea Level 500 feet 1000 feet 1500 feet 2000 feet 2500 feet 3000 feet 3500 feet 4000 feet<br />

2.00± .02 2.02± .02 2.04± .02 2.05± .02 2.08± .02 2.10± .02 2.12± .02 2.14± .02 2.16± .02<br />

5.00± .03 5.05± .03 5.10± .03 5.14± .03 5.19± .03 5.24± .03 5.29± .03 5.33± .03 5.38± .03<br />

15.00± .1 15.1± .1 15.3± .1 15.4± .1 15.5± .1 15.7± .1 15.8± .1 15.9± .1 16.1± .1<br />

25.00± .2 25.2± .2 25.4± .2 25.6± .2 25.8± .2 26.0± .2 26.2± .2 26.4± .2 26.6± .2<br />

30.00± .2 30.2± .2 30.5± .2 30.7± .2 30.9± .2 31.2± .2 31.4± .2 31.6± .2 31.9± .2<br />

Flow Flow Flow Flow Flow Flow Flow Flow Flow<br />

Sea Level 4500 feet 5000 feet 5500 feet 6000 feet 6500 feet 7000 feet 7500 feet 8000 feet<br />

2.00± .02 2.18± .02 2.20± .02 2.23± .02 2.26± .02 2.29± .02 2.31± .02 2.34± .02 2.37± .02<br />

5.00± .03 5.44± .03 5.5± .1 5.6± .1 5.6± .1 5.7± .1 5.8± .1 5.8± .1 5.9± .1<br />

15.00± .1 16.2± .1 16.4± .1 16.6± .1 16.7± .1 16.9± .1 17.1± .1 17.3± .1 17.5± .1<br />

25.00± .2 26.9± .2 27.1± .2 27.4± .2 27.7± .2 28.0± .2 28.3± .2 28.6± .2 28.9± .2<br />

30.00± .2 32.1± .2 32.4± .2 32.7± .2 33.0± .2 33.4± .2 33.8± .2 34.1± .2 34.5± .2<br />

Air and O2 Regulator Balance<br />

Objective<br />

To define the calibration requirements for balancing the Air and O2 Regulators.<br />

Equipment<br />

Pressure gauge, 0 - 30 psig, ± 1/4% F.S.<br />

Differential pressure gauge, 2 - 0 - 2 cmH2O, ± 5% F.S. Flow meter, 0 - 10 L/min, ±<br />

3% F.S.<br />

Requirements<br />

Regulator and Bleed Valve Adjustment:<br />

Set up per Standard Test Settings except set MODE to STANDBY<br />

Set O2 Blender as noted (Table 4-5)<br />

Set O2 supply pressure to 50 psi<br />

Patient circuit may be omitted<br />

L2342 Revision B November 2004