96 Bear Cub 750PSV<br />
Table 9-3: Valid Requests Received by the Ventilator<br />
Format:<br />
[ SYN ] [ID] [Check]<br />
[ SYN ]: 16h<br />
[ ID ]: 20h Send all Adult Control Settings<br />
[ Check ]: Check Sum<br />
40h Send all Adult Alarm Settings/Status<br />
60h Send all Adult Monitored Data<br />
A0h Send all Infant Control Settings<br />
C0h Send all Infant Alarm Settings/Status<br />
E0h Send all Infant Monitored Data<br />
Table 9-4: Valid Message Transmitted by the Ventilator<br />
Format:<br />
[ SYN ] [ID] [Data #1] [Data #n] [Check]<br />
[ SYN ]: 16h<br />
[ ID ]: 00h..0Fh Real Time Data, No Alarms<br />
[ Data #X ]: Binary Data<br />
[Check[: Check Sum<br />
10h 1Fh Real Time Data, Alarm Condition<br />
20h Adult Control Settings<br />
40h Adult Alarm Settings<br />
60h Adult Monitor Settings<br />
A0h Infant Control Settings<br />
C0h Infant Alarm Settings<br />
E0h Infant Monitor Settings<br />
L2342 Revision B November 2004