Service <strong>Manual</strong> 97<br />
Table 9-5: Real Time Data<br />
Mode: Continuously, every 10-msec<br />
Length: 6 Bytes<br />
Format:<br />
[ SYN ] [ID] [Data #1] [Data #3] [Check]<br />
[ ID ]: Bit 7 — 0<br />
Bit 6 — 0<br />
Bit 5 — 0<br />
Bit 4 — X<br />
Bit 3 — Y<br />
Bit 2 — Y<br />
Bit 1 — Y<br />
Bit 0 — Z<br />
Alarm Condition (x):<br />
0 — No Alarm<br />
1 — Alarm<br />
Breath Type (yyy):<br />
000 — Volume<br />
001 — Spontaneous<br />
010 — Pressure Control<br />
011 — Pressure Support<br />
Alarm Condition<br />
Breath Type<br />
Breath Phase<br />
100 — Time Cycled/Pressure Relief<br />
101 — Reserved<br />
110 — Reserved<br />
111 — Reserved<br />
Breath Phase (z):<br />
0 — Inspiration<br />
1 — Exhalation<br />
[Data #1] [Data #2] [Data #3]:<br />
xxxxxxxx xxxxxyyy yyyyyyyy<br />
| | | |<br />
MSB LSB\MSB LSB<br />
x: Net Flow (Insp-Exhale) - 13 bits<br />
Range: -300 to +300 lpm<br />
Binary Scale: 1 bit = 0.1 lpm<br />
Binary Range: 0 to 6000<br />
Zero offset: 3000<br />
y: Proximal Pressure - 11 bits<br />
Range: -60 to +140 cmH2O<br />
Binary Scale: 1 bit = 0.1 cmH2O<br />
Binary Range: 0 to 2000<br />
Zero offset: 600<br />