Service <strong>Manual</strong> 103<br />

Table 9-11: Infant Ventilator Monitor Status<br />

Mode: Upon Request<br />

Length: 20 Bytes<br />

Format:<br />

[ SYN ] [ID] [Data #1] [Data #11] [Check]<br />

[ ID ] : E0h<br />

[ Data #1 ] Breath Data<br />

Bit 7 — Control Breath (1=on, 0=off)<br />

Bit 6 — Patient Effort (1=on, 0=off)<br />

Bit 5 — Reserved (0)<br />

Bit 4 — Reserved (0)<br />

Bit 3 — Reserved (0)<br />

Bit 2 — Reserved (0)<br />

Bit 1 — Reserved (0)<br />

Bit 0 — Reserved (0)<br />

[Data #n] Parameter Range Binary Scale Binary Range<br />

[Data #2] Tidal Volume MSB 0 - 500 mL 1 mL 0 - 500<br />

[Data #3] Tidal Volume LSB " " "<br />

[Data #4] Minute Volume MSB 0 - 30.0 L 0.1 L 0 -<br />

300<br />

[Data #5] Minute Volume LSB " " "<br />

[Data #6] % Tube Leak 0 - 100% 1% 0 - 100<br />

[Data #7] Insp Time MSB 0 - 3.10 sec 0.01sec 0 - 310<br />

[Data #8] Insp Time LSB " " "<br />

[Data #9] Exp Time MSB 0 - 99.99 sec 0.01 sec 0 - 9999<br />

[Data #10] Exp Time LSB " " "<br />

[Data #11] I:E Ratio<br />

[Bit 7=Direction] 1 if 1:X.X, 0 if X.X:1<br />

[Data #12] Breath Rate 0 - 255 bpm 1 bpm 0 - 255<br />

[Data #13] Peak Insp Pres 0 - 99 cmH20 1 cmH20 0 - 99<br />

[Data #14] Mean Pres MSB 0 - 99.9 cmH20 0.1 cmH20 0 - 999<br />

[Data #15] Mean Pres LSB " " "<br />

[Data #16] Air Supply Pres 0 - 99 psig 1 psig 0 - 99<br />

[Data #17] O2 Supply Press 0 - 99 psig 1 psig 0 – 99<br />

L2342 Revision B November 2004