Technical Overview of the Cryotherapy SystemPREPARING FOR USEBefore cryosurgical treatment procedures are performed each day, thecryotherapy system must be prepared for use—according to the followingguidelines—to help ensure safe, proper functioning.Note: Before preparing the system for use, it is important to ensure thatit was appropriately processed after last use for infection prevention.For more information, see Post-Procedure Processing (page D-6).Checking the GasCylinderConnecting theRegulator to the GasCylinder• Ensure that the gas cylinder is not warm/hot to the touch beforeattaching it to the system. Do not use (or store) the cylinder nearheat sources (e.g., radiators, furnaces). Heat increases the pressureof the gas in the cylinder. This pressure can, in turn, damage thecryotherapy unit or break the rupture disk in the safety valve,causing the system to stop working. If the cylinder is warm/hot to the touch, move it to a cool placeand let it stand overnight, if possible. If the cylinder is still warm to the touch, wrap it in wet clothsuntil cool.• Connect the regulator to the gas cylinder.• Tighten the US fitting by hand only. If you are using a British fitting,which requires a wrench, tighten only until secure. Be careful not toover-tighten. (For more information about tightening the regulatorfittings, see Figure D-3.)Figure D-3. Additional Instructions for Tightening Regulator FittingsThere are two different types of fittings (Figure D-4) for connecting the regulator to the gas cylinder:• The British fitting (left) requires a separate washer. Before connecting the regulator to the cylinder via theBritish fitting, place the washer over the brass nipple that extends from the fitting.– Slide the brass connecting-nut over the nipple and hand-tighten to the cylinder connector.– Using a spanner wrench, tighten the brass nut only until secure. Be careful not to over-tighten.• The US fitting (right) does not need a separate washer, as a white plastic insert is permanently installed in theend of the nipple.– Slide the black hand-wheel with the brass nut over the nipple end and connect to the cylinder connector.– Tighten by hand only. Be careful not to over-tighten.Cervical Cancer Prevention Guidelines for Low-Resource Settings D-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!