<strong>Levelflex</strong> M <strong>FMP40</strong> with HART/4...20 mA Trouble-shooting<br />
Device displays a<br />
level when the<br />
tank is empty.<br />
Measured value<br />
incorrect (slope<br />
error in the entire<br />
measuring range)<br />
20 mA/100%<br />
actual<br />
expected<br />
4 mA/0% t →<br />
20 mA/100%<br />
L00-FMP4xxxx-19-00-00-en-020<br />
expected<br />
2.) actual<br />
1.) actual<br />
4 mA/0% t →<br />
L00-FMP4xxxx-19-00-00-en-021<br />
Incorrect probe length 1. Carry out automatic probe length<br />
detection when the tank is<br />
empty.<br />
2. Carry out mapping over entire<br />
probe when the tank is empty<br />
(probe free!).<br />
Tank properties<br />
incorrect.<br />
Medium properties<br />
incorrect.<br />
LN < 4 m and "Aluminium tank" tank<br />
properties selected<br />
→ Calibration not possible.<br />
→ Selection<br />
→ Select standard<br />
→ Thresholds too high<br />
Select lower medium properties.<br />