JUNE 26 MONDAY AFTERNOON<br />
WS-18-MoA-OR 5 DYNAMIC EXPANSION SYSTEM AT CEM. C.Matilla;<br />
N.Medina; S.Ruiz. Spanish Centre of Metrology (CEM).<br />
The Spanish Metrology Centre has an orifice-flow primary pressure standard. In this kind of pressure<br />
standard a known flow of gas passes through an orifice of calculated conductance creating a pressure<br />
gradient between both sides of the orifice. The known flow of gas is generated by a precision flowmeter<br />
which can work using the constant volume method or the constant pressure method. Since its<br />
construction the system has suffered several modifications in order to improve its results which will<br />
be described as follows.<br />
The first flowmeter was manually operated. That is the reason why it could basically be used with<br />
the constant volume method. Nowadays it has been completely automatised<br />
and it uses the constant pressure method, which is much more accurate. Data acquisition software<br />
has been designed to monitor all the parts of the system involved. Some parts of the flowmeter have<br />
been correctly isolated by means of aluminium walls to stabilise their temperature.<br />
The first vacuum pump connected to the vacuum chamber was a 300l/s turbomolecular pump. At<br />
present it has been replaced by a 700 l/s turbomolecular pump which allows the system to be closer<br />
to ideal conditions as well as to obtain an appropriate base pressure more easily.<br />