CLEANING<br />

Unit part<br />

Cuspidor <strong>unit</strong><br />

• Cleaning of Assistant’s hol<strong>de</strong>r<br />

Hol<strong>de</strong>r sleeve<br />

▪ Use the cleaner FD333 manufactured by Durr<br />

<strong>Dental</strong> AG. (Spray disinfectant mainly composed<br />

of disinfection ethanol and wipe it off, when<br />

general disinfectant is used.)<br />

▪ The Assistant’s hol<strong>de</strong>r sleeve may be <strong>de</strong>tached<br />

and sterilized in an autoclave (for 20 minutes at<br />

121°C or 12 minutes at 132°C). Remove residues<br />

and clean carefully with running water before<br />

sterilization.<br />

▪ Be sure to put into a sterilization pouch when<br />

sterilizing.<br />

▪ The hol<strong>de</strong>r sleeves are ma<strong>de</strong> of resin and may<br />

become <strong>de</strong>teriorated if they are sterilized in an<br />

autoclave many times. The allowable sterilization<br />

frequency in an autoclave is 250 times.<br />

▪ After autoclave sterilization, the hol<strong>de</strong>r sleeves<br />

are subject to discoloration, which does not have<br />

a negative effect on performance.<br />

• Assistant’s operation panel<br />

CAUTION<br />

When the surface of the operation panel is<br />

cleaned with disinfectant, etc., wipe off<br />

disinfectant completely. If it penetrates into the<br />

back of the sheet, the membrane switches may<br />

malfunction.<br />