ELECTROMAGNETIC COMPATIBILITY (EMC)<br />
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put<br />
into service according to the EMC information provi<strong>de</strong>d in this manual.<br />
Portable and mobile RF communications equipment can affect medical electrical equipment.<br />
The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or<br />
stacked use is necessary, the equipment or system should be observed to verify normal operation in the<br />
configuration in which it will be used.<br />
Guidance and manufacture’s <strong>de</strong>claration – electromagnetic emissions<br />
The CP-ONE PLUS is inten<strong>de</strong>d for use in the electromagnetic environment specified below. The customer or the<br />
user of the CP-ONE PLUS should assure that it is used in such an environment.<br />
Emissions test Compliance Electromagnetic environment - guidance<br />
RF emissions<br />
CISPR 11<br />
RF emissions<br />
CISPR 11<br />
Harmonic emissions<br />
IEC 61000-3-2<br />
Voltage fluctuations/<br />
Flicker emissions<br />
IEC 61000-3-3<br />
Group 1<br />
Class B<br />
Class A<br />
Complies<br />
The CP-ONE PLUS uses RF energy only for its internal<br />
function. Therefore, its RF emissions are very low and<br />
are not likely to cause any interference in nearby<br />
electronic equipment.<br />
The CP-ONE PLUS is suitable for use in all<br />
establishments, including domestic establishments and<br />
those directly connected to the public low-voltage<br />
power supply network that supplies buildings used for<br />
domestic purposes.<br />
Guidance and manufacture’s <strong>de</strong>claration – electromagnetic imm<strong>unit</strong>y<br />
The CP-ONE PLUS is inten<strong>de</strong>d for use in the electromagnetic environment specified below. The customer or the<br />
user of the CP-ONE PLUS should assure that it is used in such an environment.<br />
IEC 60601<br />
Electromagnetic environment -<br />
Imm<strong>unit</strong>y test<br />
Compliance level<br />
test level<br />
guidance<br />
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or<br />
discharge (ESD) ±8 kV air<br />
±8 kV air<br />
ceramic file. If floors are covered<br />
IEC 61000-4-2<br />
with synthetic material, the relative<br />
Electrical fast<br />
transient/burst<br />
IEC 61000-4-4<br />
Surge<br />
IEC 61000-4-5<br />
Voltage dips, short<br />
interruptions and<br />
voltage variations<br />
on power supply<br />
input lines<br />
IEC 61000-4-11<br />
Power frequency<br />
(50/60 Hz)<br />
magnetic field<br />
IEC 61000-4-8<br />
NOTE<br />
EMC INFORMATION<br />
humidity should be at least 30%.<br />
Mains power quality should be that<br />
of a typical commercial or hospital<br />
environment.<br />
±2 kV for power<br />
supply lines<br />
±1 kV for input/output<br />
lines<br />
±2 kV for power<br />
supply lines<br />
±1 kV for input/output<br />
lines<br />
±1 kV differential ±1 kV differential Mains power quality should be that<br />
mo<strong>de</strong><br />
mo<strong>de</strong><br />
of a typical commercial or hospital<br />
±2 kV common mo<strong>de</strong> ±2 kV common mo<strong>de</strong> environment.<br />
95% dip in U T ) of a typical commercial or hospital<br />
for 0.5 cycle<br />
for 0.5 cycle<br />
environment. If the user of the<br />
40% U T<br />
40% U T<br />
CP-ONE PLUS requires continued<br />
(60% dip in U T ) (60% dip in U T ) operation during power mains<br />
for 5 cycle<br />
for 5 cycle<br />
interruptions, it is recommen<strong>de</strong>d<br />
70% U T<br />
70% U T<br />
that the CP-ONE PLUS should be<br />
(30% dip in U T ) (30% dip in U T ) powered from an uninterruptible<br />
for 25cycle<br />
for 25cycle<br />
power supply or a battery.<br />
95% dip in U T )<br />
for 5 s<br />
for 5 s<br />
3 A/m 3 A/m Power frequency magnetic fields<br />
should be at levels characteristic of<br />
a typical location in a typical<br />
commercial or hospital<br />
environment.<br />
U T is the a.c. mains voltage prior to applications of the test level.<br />
77