Unit part<br />
CLEANING<br />
Doctor’s <strong>unit</strong><br />
Oil mist<br />
separator<br />
Operation<br />
panel<br />
Tray mat<br />
Hol<strong>de</strong>r on doctor si<strong>de</strong><br />
Dirt bin<br />
• Dirt bin<br />
(2)<br />
(1)<br />
(3)<br />
▪ If dirt is collected, remove the paper cup (2) from<br />
the dirt bin hol<strong>de</strong>r (1) and replace it.<br />
▪ The stainless dirt bin (3) (optional) may be<br />
<strong>de</strong>tached when it is turned in direction A. It is<br />
fastened when turned in direction B.<br />
The cover has sharp portions that can easily catch<br />
cotton, etc. Be very careful when cleaning it.<br />
• Oil mist separator<br />
(1)<br />
(2)<br />
▪ The oil mist separator collects oil contained in<br />
exhaust air from the handpieces.<br />
▪ Be sure to dispose of oil when oil is collected up<br />
to the red line of the oil reservoir (1).<br />
▪ The oil reservoir may be <strong>de</strong>tached when it is<br />
rotated counterclockwise.<br />
▪ Replace the moisture absorbing sponge (2) if a<br />
lot of dirt or oil adheres to it.<br />
• Tray mat<br />
▪ The tray mat may be <strong>de</strong>tached and sterilized in an<br />
autoclave (for 20 minutes at 121°C or 12 minutes<br />
at 132°C). Remove residues and clean carefully<br />
with running water before sterilization.<br />
▪ Be sure to put into a sterilization pouch when<br />
sterilizing.<br />
▪ The tray mat is ma<strong>de</strong> of resin and may become<br />
<strong>de</strong>teriorated if it is sterilized in an autoclave<br />
many times.<br />
The allowable sterilization frequency in an<br />
autoclave is 250 times.<br />
▪ After autoclave sterilization, the hol<strong>de</strong>r sleeves<br />
are subject to discoloration, which does not have<br />
a negative effect on performance.<br />
67