Unit part<br />

CLEANING<br />

Utility box<br />

• Cleaning air filter drain valve and discharging water from air compressor<br />

Drain valve<br />

▪ This valve is used to discharge water from the air<br />

filter.<br />

▪ Turn the drain valve counterclockwise to discharge<br />

water collected in the air filter once a week at least.<br />

▪ If water enters the <strong>unit</strong>, the air turbine, air motor or<br />

syringe, etc., may become <strong>de</strong>fective. Be sure to turn<br />

the drain valve clockwise to close the valve after<br />

discharging water from the air filter.<br />

▪ Open the drain valve of the air compressor to<br />

discharge collected water once a week.<br />

Outsi<strong>de</strong> of the product<br />

• Cleaning outsi<strong>de</strong> of the product<br />

▪ Clean the metallic parts with a dry soft cloth.<br />

Wipe off water immediately if water is put on the product. Water may cause rusting.<br />

▪ Clean the resin parts with a wet soft cloth.<br />

• Disinfecting outsi<strong>de</strong> of the product<br />

▪ Use FD333 manufactured by Durr <strong>Dental</strong> AG to clean and disinfect the outsi<strong>de</strong> of the product.<br />

NOTICE<br />

Wipe off remaining water and disinfectant immediately. They may facilitate rust generation.<br />