Unit part<br />

CLEANING<br />

Doctor’s <strong>unit</strong><br />

• 77 type 3-way syringe<br />

Nut<br />

Nozzle<br />

▪ The nozzle may be sterilized in an autoclave (for 20<br />

minutes at 121°C or 12 minutes at 132°C). Remove<br />

residues and clean carefully with running water before<br />

sterilization.<br />

▪ Be sure to put into a sterilization pouch when sterilizing. The<br />

allowable sterilization frequency in an autoclave is 250 times.<br />

(Standard of handpiece for <strong>de</strong>ntal use JIS T 5906:01)<br />

The nut may be removed from the syringe by turning it in<br />

direction A. Apply Vaseline thinly and evenly to the two<br />

O-rings after sterilization. Tighten the nut firmly in direction B<br />

to undo the nozzle.<br />

Before use, make sure that the nut is firmly tightened.<br />

CAUTION<br />

• RADIS (BT) 3-way syringe<br />

(3)<br />

(1)<br />

(2)<br />

(4)<br />

▪ Nozzle (1) and body cover (2) may be sterilized in an<br />

autoclave (for 20 minutes at 121°C or 12 minutes at 132°C).<br />

Remove residues and clean carefully with running water<br />

before sterilization.<br />

▪ Be sure to put into a sterilization pouch when sterilizing.<br />

▪ The body cover is ma<strong>de</strong> of resin and may become<br />

<strong>de</strong>teriorated if it is sterilized in an autoclave many times.<br />

The allowable sterilization frequency in an autoclave is<br />

250 times.<br />

(Standard of handpiece for <strong>de</strong>ntal use JIS T 5906:01)<br />

▪ After autoclave sterilization, the body cover is subject to<br />

discoloration, which does not have a negative effect on<br />

performance.<br />

▪ The nozzle may be <strong>de</strong>tached by turning the nut (3) in<br />

direction A. The body cover may be <strong>de</strong>tached by pulling it in<br />

direction C while pressing the button (4).<br />

▪ Apply Vaseline thinly and evenly to two O-rings of the<br />

nozzle after sterilization and turn the nut (3) firmly in<br />

direction B.<br />

Before use, make sure that the nut is firmly tightened.<br />

CAUTION<br />

NOTICE<br />

For operation of other syringes, observe the instructions shown in the package inserts and Instruction<br />

Manuals inclu<strong>de</strong>d with the syringes.<br />