<strong>CCOS</strong> Field Operations Plan Version 2: 5/31/00<br />

Device Manufacturer Model Element Range<br />

Bubble meter Teledyne/Hastings 0 - 5 lpm<br />

Laminar flow meter Meriam 50MK10 725250-C1 0 - 50 cc/min<br />

Laminar flow meter Meriam 50MK10 725250-C2 0 - 1 lpm<br />

Laminar flow meter Meriam 50MJ10-1/2 700310-T2 0 - 10 lpm<br />

Laminar flow meter Meriam 50MJ10-1/2 700310-T1 0 - 50 lpm<br />

Laminar flow meter Meriam 50MW20-1 700310-T3 0 - 200 lpm<br />

The Laminar flow devices are certified annually against a primary flow standard by an external<br />

laboratory. The bubble flow meter is a primary flow measuring device, and therefore requires no<br />

certification. Parsons also maintains a bubble flow meter.<br />

Flow Meters - Several flow meters are available as transfer standards for this study.<br />

These flow meters and their operating ranges are summarized below:<br />

Make Model SN Range Type<br />

Bubble-O 0 - 400 ccpm Bubble<br />

Bios DC-Lite 0 - 20 lpm Piston<br />

Schlumberger Remus 4, G1-6 0 - 20 lpm Dry test<br />

Teledyne-Hastings HS-10KS 16045 0 - 10 lpm Mass flow<br />

Teledyne-Hastings HS-50JS 16944 0 - 50 lpm Mass flow<br />

Teledyne-Hastings HS-10S 17259 0 - 150 lpm Mass flow<br />

All flow meters will be certified against suitable flow standards prior to being used in the field.<br />

Certifications consist of a comparison between the flow meter and the appropriate QA laminar<br />

flow standard at at least 10 different flow spaced evenly across the operating range of the mass<br />

flow meter.<br />

Continuous Air Quality Analyzers<br />

Dilution Standards and Gases - Multipoint audits will be performed using either a Dasibi<br />

1009 CP dilution system, an Environics Series 100 dilution system, or equivalent system. These<br />

systems are maintained and the flow rates certified by Parsons’ and the<br />

departments. The zero air source will be either a cylinder of Ultra-Pure Air that will be supplied<br />

by Scott Specialty Gases or an Aadco 737R pure air system with CH4 burner and compressor<br />

capable of delivering a constant 20 lpm air supply. The span gases, except for the SUVA<br />

refrigerant, will be provided by Scott Specialty Gases. All gas standard analysis is performed in<br />

accordance with EPA Protocols and is traceable to National <strong>Institute</strong> of Standards and<br />

Technology (NIST) standards.) For the <strong>CCOS</strong> audits, Relative Accuracy Test Audits (RATA)<br />

Class cylinders will be used, which are certified to with ±1% of true. SUVA refrigerant will be<br />

used to audit the nephelometers. This compound has a known scattering coefficient that is<br />

within the nephelometer measurement range.<br />

<strong>Ozone</strong> - All field ozone transfer standards will be certified against a primary standard.<br />

Procedures for certification and recertification of ozone primary and transfer standards are based<br />

on the EPA technical assistance document, Transfer Standards for Calibration of Air Monitoring<br />

Analyzers for <strong>Ozone</strong> (EPA 1979), EPA-600/4-79-056. All ozone transfer standards used for<br />

Chapter 4: QUALITY ASSURANCE 4-8