<strong>CCOS</strong> Field Operations Plan (DRAFT) Version 2: 5/31/00<br />

Table 2.4-4<br />

Primary, calibration, performance test and performance audit standards and frequencies for <strong>CCOS</strong> measurements<br />

Observable<br />

Expected<br />

Precision<br />

Primary Standard<br />

O3 by ultraviolet absorption ±10% Dasibi 1008PC UV<br />

photometer,<br />

referenced to ARB<br />

primary<br />

NO/NOY by<br />

chemiluminescence<br />

NOY/NOY* by<br />

chemiluminescence<br />

NO/NOX by<br />

chemiluminescence<br />

CO by infrared gas filter<br />

correlation<br />

Calibration<br />

Standard<br />

Dasibi 1003PC UV<br />

photometer<br />

±10% NIST-traceable NO Certified NO w/<br />

dynamic dilution and<br />

gas-phase titration<br />

±10% NIST-traceable NO Certified NO w/<br />

dynamic dilution and<br />

gas-phase titration<br />

±10% NIST-traceable NO Certified NO w/<br />

dynamic dilution and<br />

gas-phase titration<br />

±10% NIST-traceable CO Certified CO w/<br />

dynamic dilution<br />

Calibration<br />

Frequency<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

Performance Test Standard<br />

Span and precision check with calibrated<br />

ozone generator and zero with scrubbed<br />

air<br />

Span and precision check with certified<br />

NO, converter check with NO2 by gas<br />

phase titration, and zero with scrubbed air<br />

Span and precision check with certified<br />

NO, converter check with NO2 by gas<br />

phase titration, and zero with scrubbed air<br />

Span and precision check with certified<br />

NO, converter check with NO2 by gas<br />

phase titration, and zero with scrubbed air<br />

Span and precision check with certified<br />

CO and zero with scrubbed air<br />

Performance<br />

Test<br />

Frequency<br />

Daily<br />

Performance Audit<br />

Standard<br />

Dasibi 1008PC with<br />

temperature and<br />

pressure adjustments<br />

Daily Certified NO w/<br />

dynamic dilution and<br />

NO2 by gas-phase<br />

titration<br />

Daily Certified NO w/<br />

dynamic dilution and<br />

NO2 by gas-phase<br />

titration<br />

Daily Certified NO w/<br />

dynamic dilution and<br />

NO2 by gas-phase<br />

titration<br />

Daily Certified CO w/<br />

dynamic dilution<br />

Performance<br />

Audit<br />

Frequency<br />

Once during<br />

study<br />

Once during<br />

study<br />

Once during<br />

study<br />

Once during<br />

study<br />

Once during<br />

study<br />

CO2 by infrared gas filter<br />

correlation<br />

±10% NIST-traceable CO2 Certified CO2 w/<br />

dynamic dilution<br />

At start and end of<br />

study<br />

Span check at two levels with certified<br />

CO2 and zero with scrubbed air<br />

Daily - -<br />

NO2, PAcN by<br />

chemiluminescence and<br />

luminol<br />

±20% NIST-traceable NO2<br />

and PAN<br />

HCHO by methanalyzer ±20% NIST-traceable<br />

HCHO<br />

PM2.5 nitrate by flash<br />

volatilization<br />

Light scattering by<br />

nephelometers<br />

Light absorption by<br />

aethalometers<br />

Certified NO w/<br />

dynamic dilution and<br />

gas-phase titration<br />

Certified HCHO<br />

permeation tube<br />

±15% NIST-traceable NO Certified NO w/<br />

dynamic dilution and<br />

nitrate solution<br />

±10% HFC-134a refrigerant HFC-134a, pure<br />

CO2, and particle<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

At start and end of<br />

study<br />

Span check with NO2 by gas phase<br />

titration and zero with scrubbed air<br />

Internal zero check, HCHO permeation<br />

tube<br />

Span with certified NO and zero with<br />

scrubbed air<br />

At start and end of<br />

study<br />

free air<br />

±5 to 20% Neutral density filters Factory calibration When out of spec Internal electronic checks and neutral<br />

density filters<br />

Daily for NO2 Certified NO2 w/<br />

dynamic dilution and<br />

gas-phase titration<br />

Zero: once/10<br />

minutes,<br />

HCHO at end<br />

of study<br />

Weekly Certified NO w/<br />

dynamic dilution and<br />

nitrate solution<br />

Once during<br />

study<br />

- -<br />

HFC-134a, and particle free air Weekly HFC-134a, pure<br />

CO2, and particle<br />

free air<br />

Weekly and<br />

monthly<br />

Neutral density filters<br />

Once during<br />

study<br />

Once during<br />

study<br />

Once during<br />

study<br />

Chapter 2: FIELD MEASUREMENTS 2-28