C.6 Supplemental Measurements of Volatile Organic Compounds<br />
A.7.1 Collection and Analysis of Hydrocarbons and Oxygenated Species<br />
Canister Sampling<br />
Stainless steel SUMMA-polished canisters of 6-L capacity are customarily employed<br />
for volatile hydrocarbon (C 2 -C 12 ) collection. These canister samples are to be analyzed for<br />
speciated hydrocarbons by a method consistent with EPA Method TO-14, as well as for CO, CO 2 ,<br />
methane, and oxygenated species. Prior to sampling, the canisters are to be cleaned by repeated<br />
evacuation and pressurization with humidified zero air, and certified as described in the EPA<br />
document "Technical Assistance Document for Sampling and Analysis of <strong>Ozone</strong> Precursors"<br />
(October 1991, EPA/600-8-91/215).<br />
The sampling procedure should essentially follow the pressurized sampling method<br />
described by EPA Methods TO-12 and TO-14 and the EPA document "Technical Assistance<br />
Document for Sampling and Analysis of <strong>Ozone</strong> Precursors" (October 1991, EPA/600-8-91/215).<br />
A stainless steel Viton pump draws in ambient air from the sampling manifold to fill and<br />
pressurize the sample canisters. A flow control device maintains a constant flow into the canisters<br />
over the desired sample period. This flow rate is preset to fill the canisters to about 1 atm above<br />
ambient pressure at the end of the sampling period (as described by EPA Method TO-14). For<br />
automatic operation, the timer starts and stops the pump at the appropriate time. The timer also<br />
opens the solenoid valve when the pump starts and closes it when the pump stops. The use of the<br />
solenoid manifold valves permits the automatic selection of preloaded canisters. The multipleevent<br />
sampling systems, allowing unattended collection of up to six canister samples are<br />
recommended for this study.<br />
After sampling, an identification tag should be attached to each canister with the canister<br />
serial number, sample number, and sampling location, date, and time recorded on this tag. In<br />
addition a field sampling form and chain-of-custody form should be filled out giving all pertinent<br />
information on the collection of the sample.<br />
GC/FID Analysis<br />
Gas chromatography with flame ionization detector is the established technique for<br />
monitoring volatile hydrocarbons, ozone precursors, in ambient air. An air sample is taken from<br />
the canister and passed through the sample concentration system. This system usually consists of a<br />
freeze-out loop, made from chromatographic-grade stainless steel tubing packed with 60/80 mesh<br />
deactivated glass beads, and a 10-port sampling valve. First, the sample is transferred from the<br />
canister through the loop immersed in liquid oxygen to the volume transfer measurement apparatus.<br />
The C 2 and heavier hydrocarbons are cryogenically trapped inside the loop when air is transferred<br />
to an evacuated flask of known volume. From the difference in pressure inside the flask, the<br />
volume of the air sample can be calculated, based on the ideal gas law. When a sufficient volume<br />
of the air sample has been transferred from the canister to the concentration system, the 10-port<br />
valve is switched to position 2, the liquid oxygen is replaced with boiling water, and the contents<br />
of the trap are injected into a chromatographic column where separation of the C 2 -C 12<br />
C-12