Appendix A: Specifications<br />

Table A- 10: Certifications and compliances<br />

Category<br />

EC Declaration of Conformity -<br />

EMC<br />

Standards or description<br />

Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was<br />

demonstrated to the following specifications as listed in the Official Journal of the European Union:<br />

EN 61326 Emissions 1, 3, 4 Class A Radiated and Conducted Emissions<br />

EN 61326 Immunity 1, 4<br />

IEC 61000-4-2 Electrostatic Discharge Immunity<br />

±4 kV contact discharge, ±8 kV air discharge<br />

IEC 61000-4-3 RF field immunity<br />

3V/m,80MHzto1GHz,<br />

80% amplitude modulated with a 1 kHz sinewave 2<br />

IEC 61000-4-4 Electrical Fast Transient/Burst Immunity<br />

1 kV on AC mains, 500 V on I/O<br />

IEC 61000-4-5 AC Surge Immunity<br />

500 V differential mode, 1 kV common mode<br />

IEC 61000-4-6 RF Conducted Immunity<br />

3 V, 150 kHz to 80 MHz, amplitude modulated with a<br />

1 kHz sinewave 2<br />

IEC 61000-4-11 AC Mains Voltage Dips and Interruption Immunity<br />

100% reduction for one cycle<br />

EN 61000-3-2<br />

Power Harmonic Current Emissions<br />

EN 61000-3-3<br />

Voltage Changes, Fluctuations, and Flicker<br />

1 Use low-EMI shielded interconnect cables, equivalent to the following Tektronix cables:<br />

GPIB cable: 012-0991-01, 012-0991-02, or 012-0991-03.<br />

RS-232 cable: 012-1213-00 or CA part number 0294-9.<br />

Centronics Cable: 012-1214-00 or LCOM part number CTL3VGAMM-5 VGA Cable.<br />

2 Under theses conditions, the specifications are amended as follows:<br />

1 mV/division to 1 V/division: ≤0.2 division waveform displacement or ≤0.4 division increase<br />

in peak-to-peak noise.<br />

3 Radiated emissions may exceed the levels specified in EN 61326 when this instrument is<br />

connected to a test object.<br />

4 Tested in accordance with EN 61326 Annex D.<br />

FCC Radiated and conducted emissions do not exceed the levels specified in FCC47 CFR, Part 15,<br />

Subpart B, for Class A equipment.<br />