Appendix B: Performance Verification<br />

b. Connect the other cable end to the digital multimeter with a BNC ‘T’<br />

connector, 75 Ω precision termination, and BNC female-to-dual banana<br />

adapter.<br />

DMM<br />

TG700 with<br />

AWVG7 module<br />

BNC to Dual banana<br />

75 Ω Termination<br />

CH3 CH2 CH1<br />

BNC “T”<br />

connector<br />

Figure B- 1: Signal output amplitude test hookup<br />

2. On the TG700, press the MODULE, FORMAT, andFRONT PANEL<br />

ENABLE buttons simultaneously, and then release the MODULE and<br />

FORMAT buttons. Continue to hold the FRONT PANEL ENABLE button<br />

to display Factory Startup Mode.<br />

3. Enable the DAC Gain (GBR) : 0 mV calibration signal as follows:<br />

a. Press the MODULE button to display the AWVG7 main menu.<br />

b. Press the up or down arrow button to select CALIBRATION.<br />

c. Press the left or right arrow button to select DAC Gain (GBR).<br />

d. Press the ENTER button.<br />

e. Press the left or right arrow button to select 0mV.<br />

f. Press the ENTER button.<br />

4. Enter the digital multimeter reading in Table B -2, row DC0, column CH1.<br />

5. Move the BNC cable from the upper CH1 connector to the upper CH 2<br />

connector on the AWVG7 generator module.<br />

6. Enter the digital multimeter reading in Table B -2, row DC0, column CH 2.<br />

VM5000 Automatic Video Measurement Set <strong>User</strong> <strong>Manual</strong> B-3