Read This First<br />

Type N to BNC adapter (Qty 2)<br />

<br />

<br />

75 Ω BNC female/female adapter<br />

50 Ω BNC female/female adapter<br />

1. Connect the MIU to the VM5000, apply power, and establish communications<br />

using Hyperterminal as described in Test A of this procedure.<br />

2. Check that the MIU RGB WB Path LED is lit. If not, using Hyperterminal,<br />

send the command $LOAD=1000000 to the MIU. Verify that the RGB WB<br />

Path LED is illuminated.<br />

3. On the VM5000 front panel, under VERTICAL, select 50 Ω for CH1, CH2,<br />

and CH3.<br />

4. Set the tracking generator and spectrum analyzer controls as follows:<br />

<br />

<br />

<br />

<br />

<br />

<br />

Tracking Generator output level = 0 dBm<br />

Spectrum Analyzer Vertical scale = 1 dB/div<br />

Spectrum Analyzer Reference Level = -10 dBm<br />

Horizontal scale (Span) = Max Span (or 200 MHz/div if available)<br />

Center Frequency = 900 -1000 MHz (10 div display)<br />

Resolution BW (RBW) = 3 MHz<br />

5. Connect a Type N to BNC adapter, followed by a 6 dB attenuator, to the<br />

tracking generator output. Connect a 50 Ω BNC cable to the other end of the<br />

attenuator. Connect a second Type N to BNC adapter and 6 dB attenuator to<br />

the spectrum analyzer input, and connect the other end of this attenuator to a<br />

second 50 Ω BNC cable. Connect the two cables together with a 50 Ω BNC<br />

female/female adapter.<br />

6. Adjust the analyzer Reference Level control until the displayed signal is<br />

entirely on screen over the frequency range 1 MHz to 1800 MHz.<br />

7. Using the analyzer Display memory, store the swept display. Activate the<br />

“active display minus stored display” mode (sometimes called “B minus<br />

Save A”). A straight line display, with less than 0.2 dB ripple, should appear,<br />

typically at mid-screen. This is the frequency response with the flatness<br />

errors of the tracking generator, attenuators, spectrum analyzer, and cables<br />

normalized out.<br />

8. Connect two 75/50 min-loss pads together, using a 75 Ω BNC female/female<br />

adapter. Insert this connection in place of the 50 Ω BNC female/female<br />

adapter. Decrease the analyzer Reference Level by 11 dB, so that the<br />

frequency response of the series connected min-loss pads can be observed.<br />

Measure the response at 1500 MHz, relative to 1 MHz, and record this value<br />

16 VM5000 Automatic Video Measurement Set Option VGA Read This First