Appendix B: Performance Verification<br />

8-bit INL, DNL, Monotonicity, and Resolution measurement. The steps for checking<br />

the 8-bit INL, DNL, Monotonicity, and Resolution measurement are as follows:<br />

11. From the VM5000 application menu bar, click the Run button to take the<br />

linearity measurement.<br />

12. Record the measured result in Table B -17.<br />

13. Check that the results are within the upper and lower limit.<br />

Table B- 17: 8-bit Linearity measurement results - G channel<br />

Linearity Lower limit G Upper limit<br />

Resolution (Bits) 8 8<br />

Max INL (LSB) -5.12 5.12<br />

Max DNL (LSB) -0.25 0.25<br />

Monotonic (Y/N) Yes Yes<br />

14. Connect the AWG710B CH1 output to the B input of the MIU.<br />

15. From the VM5000 application menu bar, click the Run button to take the<br />

linearity measurement.<br />

16. Record the measured result in Table B -18.<br />

17. Check that the results are within the upper and lower limit.<br />

Table B- 18: 8-bit Linearity measurement results - B channel<br />

Linearity Lower limit B Upper limit<br />

Resolution (Bits) 8 8<br />

Max INL (LSB) -5.12 5.12<br />

Max DNL (LSB) -0.25 0.25<br />

Monotonic (Y/N) Yes Yes<br />

18. Connect the AWG710B CH1 output to the R input of the MIU.<br />

19. From the VM5000 application menu bar, click the Run button to take the<br />

linearity measurement.<br />

20. Record the measured result in Table B -19.<br />

21. Check that the results are within the upper and lower limit.<br />