Read This First<br />

Test A: Initial Connection and Setup<br />

1. With the VM5000 Power off, connect all five MIU coaxial cables to the host<br />

VM5000. Use the labels on the MIU as a guide. Connect the RS232 cable<br />

between the MIU and COM1 port on the VM5000. Attach a keyboard and<br />

mouse to the VM5000. Turn on the VM5000. The MIU Power ON led<br />

should be lit and the MIU Fault LED should be off. If the Fault indicator is<br />

lit, check that all of the coaxial cables have been properly connected.<br />

2. Open Hyperterminal on the VM5000. If an “MIU – Hyperterminal” session<br />

was previously established and saved, open it. If not, establish a session as<br />

follows:<br />

a. Minimize open programs, such as the oscilloscope application and the<br />

VM5000 application, so that the Windows desktop is visible.<br />

b. Open the terminal emulation program Hyperterminal. Hyperterminal is<br />

typically located under Start>Programs>Accessories>Communications>Hyperterminal.<br />

c. When Hyperterminal is first opened, a pop-up window Connection<br />

Description will open. Under name, enter MIU, and then click OK.<br />

d. A Connect To window will open. Under Connect using, select COM1,<br />

and click OK.<br />

e. A COM1 Properties window will open. Set the Port Settings as follows:<br />

<br />

<br />

<br />

<br />

<br />

Bits per second=9600<br />

Data bits=8<br />

parity=None<br />

Stop bits=1<br />

Flow Control=None<br />

After changing the settings, click OK.<br />

3. The Hyperterminal window should be open, as connection MIU. Select<br />

File>Properties. The MIU Properties window will appear. Select the Settings<br />

tab, followed by clicking the ASCII Setup button. Select Echo typed<br />

characters locally and Send line ends with line feeds. Click OK to close the<br />

ASCII Setup window, and then click OK to close the MIU Properties<br />

window.<br />

4. From the keyboard, type $ID? The MIU should identify itself, returning a<br />

response similar to id=miu_v1.0. If there is no response, double check all<br />

electrical connections and the Hyperterminal settings described above.<br />

4 VM5000 Automatic Video Measurement Set Option VGA Read This First