Incoming Inspection<br />

This section contains instructions for performing a self test and functional<br />

verification of the instrument. These procedures verify that the instrument is<br />

operating correctly after shipment, but they do not check product specifications.<br />

If the instrument fails any test within this section, it may need service. To contact<br />

Tektronix for service, see Contacting Tektronix on the copyright page.<br />

Self Tests<br />

Make sure you have installed the instrument as described in Installation starting<br />

on page 1 -5.<br />

This procedure uses internal routines to verify that the instrument functions and<br />

was adjusted properly. No test equipment or hookups are required.<br />

Equipment required<br />

Prerequisites<br />

None<br />

Power on the instrument and allow a 20 minute warm-up.<br />

Perform these substeps to verify internal diagnostics:<br />

1. Display the system diagnostics menu as follows:<br />

a. If the instrument is in the toolbar mode, click Menu (far right toolbar<br />

button) to switch to menu bar mode.<br />

b. From the Utilities menu, select Instrument Diagnostics to open the<br />

Instrument Diagnostics control window.<br />

c. In the Subsystem, Area, andTest lists, ensure that you select ALL.<br />

2. Run the system diagnostics as follows:<br />

a. Disconnect all input signals from all four channels.<br />

b. Click Run in the Instrument Diagnostics control window.<br />

The internal diagnostics do an exhaustive verification of proper instrument<br />

function, taking five to fifteen minutes. When complete, the resulting status<br />

appears in the diagnostics control window.<br />

3. Verify that no failures are found and reported: The instrument should pass<br />

all tests. If any failures occur, record the failure information and contact your<br />

local Tektronix service personnel for more information.<br />