14. Remove the BNC short and reconnect the MIU CH2 coaxial cable to the<br />

VM5000 CH2 input.<br />

15. Connect the BNC “T” adapter to the R input BNC on the MIU.<br />

16. Disconnect the MIU CH3 coaxial cable from the VM5000 and connect the<br />

BNC short to the cable end, just as was done for CH1 (and described in<br />

step 6 above).<br />

17. Check that the green LED in the RGB LF Path is still illuminated. If not,<br />

using Hyperterminal, send the command $LOAD=0000000 to the MIU.<br />

Verify that the LED in the RGB LF Path is now illuminated.<br />

NOTE. The V channel and H/V LF Path LEDs will also be illuminated.<br />

18. CHECK – that the Multimeter reads less than 15.0Ω.<br />

NOTE. The multimeter will typically read between 10 Ω and 11 Ω. If the reading<br />

is around 75Ω, check that the BNC short is fully inserted into the MIU CH3<br />

coaxial cable BNC and that the RGB LF Path LED is illuminated.<br />

19. Remove the BNC short and reconnect the MIU CH3 coaxial cable to the<br />

VM5000 CH3 input.<br />

20. Connect the BNC “T” adapter to the V input BNC on the MIU.<br />

21. Disconnect the MIU CH4 coaxial cable from the VM5000 and connect the<br />

BNC short to the cable end, just as was done for CH1 (and described in step<br />

6 above). All of the LEDs on the MIU will be extinguished, but the MIU is<br />

still powered for the H and V channel measurements.<br />

22. Using Hyperterminal, send the command $LOAD=0000000 to the MIU. No<br />

LEDs will be illuminated, but the V channel and HV LF Path are now<br />

selected.<br />

23. CHECK – that the Multimeter reads less than 19.0 Ω.<br />

NOTE. The multimeter will typically read between 10 Ω and 11 Ω. If the reading<br />

is >100 Ω, check that the BNC short is fully inserted into the MIU CH4 coaxial<br />

cable BNC. If the >100 Ω reading persists, check that the MIU is in the proper<br />

mode by sending the query (via Hyperterminal) $LOAD?. You should receive a<br />

reply load=0000000.<br />

24. Connect the BNC “T” adapter to the H input BNC on the MIU.<br />

8 VM5000 Automatic Video Measurement Set Option VGA Read This First