Appendix B: Performance Verification<br />

Table B- 22: 10-bit Linearity, G Channel, Level 3 of 4<br />

Linearity Lower limit G Upper limit<br />

Resolution (Bits) 8 8<br />

Max INL (LSB) -20.48 20.48<br />

Max DNL (LSB) -0.5 0.5<br />

Monotonic (Y/N) Yes Yes<br />

41. From the TekScope application menu bar, select Top menu: Vertical > Offset<br />

> Select Ch1 > enter -0.175 (Volts) in Offset box.<br />

42. From the TekScope application menu bar, click on App software button.<br />

43. From the VM5000 application menu bar, click the Run button to take the<br />

linearity measurement.<br />

44. Record the results in Table B -23.<br />

45. Check that the results are within the upper and lower limits (Resolution<br />

should read 8-bit).<br />

Table B- 23: 10-bit Linearity, G Channel, Level 4 of 4<br />

Linearity Lower limit G Upper limit<br />

Resolution (Bits) 8 8<br />

Max INL (LSB) -5.12 5.12<br />

Max DNL (LSB) -0.5 0.5<br />

Monotonic (Y/N) Yes Yes<br />

46. Compute composite resolution: If the Resolution (bits) value in Tables B -20<br />

through B -23 is the same, add 2 to the value and enter this value under<br />

“Resolution (Bits)” in Table B -24. If the value for Resolution (bits) is not<br />

the same for all tables, the Resolution portion of the performance verification<br />

for this channel has failed.<br />

47. Determine maximum INL: find the worst case (extreme) INL value from<br />

Table B -20 to Table B -23 and enter the resulting value, or this value plus<br />

the INL value from Table B -17, which ever has the largest absolute value,<br />

under ”Max INL (LSB)” in Table B -24 on page B -39.<br />

48. Compute maximum DNL: find the worst case (extreme) DNL value from<br />

Table B -20 to Table B -23 and enter the resulting value under “Max DNL<br />

(LSB)” in Table B -24 on page B -39.<br />