Figure 4.1. Bitex main panel when fault came<br />

At the time when servicing started, the reason why these were displaying red and yellow<br />

were not known. So, the trouble–shooting was done in a systematic way as follows:<br />

(1) Zauto was opened and checked autocal. The calibration was ok.<br />

(2) Antenna and ascope were opened.<br />

(3) DCU general is displaying red and so, on right-clicking it, RF generator status is displayed<br />

with fault at receiver protect.<br />

(4) Then all the orange cable wiring and cable to receiver protector (J4) was checked. It was<br />

found ok.<br />

(5) Later, connected to rvp8 by ssh command and rebooted the rvp8.<br />

(6) As DAU bytes 0-3 was red, it was opened, and it was showing fault in Klystron preheat.<br />

Hence, it was concluded that preheat was not available due to receiver protect signal not<br />

available. So, it was concluded that the cables connecting to receiver protector should be<br />

checked physically.<br />

(7) Hence, all cables connected to receiver protector were checked with CRO and also<br />

physically. It was finally found that one cable no.4 connecting to receiver protector was<br />

found cut and it was soldered. <strong>Radar</strong> was turned on and then preheat became available.<br />

(8) Then HV was switched on and then radar started working ok.<br />