SA9-3<br />

position 3 of A1SA9.<br />

No. Position Parameter<br />

8 A1P5/<br />

SA9-4<br />

Table C.2. Location and Calibration of panel meters<br />

Calibration method<br />

+28 VDC Adjust A1A2R30 to make the reading of A1P5 equal to 28 volts at<br />

position 4 of A1SA9.<br />

9 A1P5/<br />

SA9-5<br />

10 A1P5/<br />

SA9-6<br />

+40 VDC Adjust A1A2R31 to make the reading of A1P5 equal to 40 volts at<br />

position 5 of A1SA9.<br />

+510 VDC Adjust A1A2R23 to make the reading of A1P5 equal to 510 volts at<br />

position 6 of A1SA9.<br />

11 A1P5/<br />

SA9-7<br />

12 A1P5/<br />

SA9-8<br />

13 A1P5/<br />

SA9-9<br />

14 A1P5/<br />

SA9-10<br />

15 A1P5/<br />

SA9-11<br />

16 A1P5/<br />

SA9-12<br />

Filament<br />

Power<br />

Supply<br />

Inverse<br />

Voltage<br />

Filament<br />

Voltage<br />

Focus Coil<br />

Voltage<br />

Ion Pump<br />

Voltage<br />

Beam<br />

Current<br />

Electron-<br />

Beam<br />

voltage<br />

Measure the peak-to-peak voltage between terminals 1 and 2 of the<br />

terminal board A7A1XT1 using an oscilloscope and read the value.<br />

Adjust A1A2R24 to make the reading of A1P5 equal to the value at<br />

position 7 of A1SA9<br />

Adjust A1A2R32 to make the reading of A1P5 equal to the nameplate<br />

filament voltage value of klystron V1 at position 8 of A1SA9.<br />

Measure the voltage value between terminals 1 and 2 of the terminal<br />

board L1XT1 using a multi-meter. Adjust A1A2R25 to make the<br />

reading of A1P5 equal to the value at position 9 of A1SA9<br />

Adjust A1A2R26 to make the reading of A1P5 equal to 3KV at position<br />

10 of A1SA9.<br />

Adjust A1A2R19 to make the reading of the A1P5 equal to 24<br />

milliamperes at position 11 of A1SA9 with the PRF 322Hz and narrow<br />

pulse.<br />

Adjust A1A2R16 and A1A2R16B to make the reading of A1P5 equal to<br />

60KV at position 12 of A1SA9.<br />

17 A1P5/<br />

SA9-13<br />

Inverse Peak Adjust A1A2R18 to make the reading of the A1P5 equal to 11<br />

Current milliamperes at position 13 of A1SA9 with the PRF 322Hz and narrow<br />

pulse.<br />

18 A1P5/<br />

SA9-14<br />

Charge<br />

Current<br />

Adjust A1A2R17 to make the reading of A1P5 equal to 0.65 amperes at<br />

position 14 of A1SA9 with the PRF 322Hz and narrow pulse.<br />

19 A1P5/ Regulating Adjust A1A2R11 to make the reading of A1P5 equal to 5 milliamperes<br />

74