APPENDIX C<br />
C.1. Calibration procedures of panel meters<br />
Control panel meters are to be calibrated. Meter A1P1 indicates filament current. Similar meter<br />
is used for focus coil current. Meter P3 is for indication of Ion pump current.P5 indicates the<br />
charging voltage. Meter P4 is used for series of 15 parameters by means of a selection switch.<br />
The following procedure is to be followed to calibrate the above meters as given in C.1 below.<br />
Figure C.1. Tx Control Panel A1, Panel Meters, Location of Controls and Indicators<br />
1. Set the H.V.Power Circuit breaker Q1, Auxiliary circuit breaker Q2, and cabinet light<br />
power circuit breaker Q3 off.<br />
2. Open the cabinet doors.<br />
3. Remove the charging switch module, A10 from TX bay.<br />
4. Close the right-bay inner door. Set the cabinet light power circuit breaker Q3 and Auxiliary<br />
circuit breaker Q2, and Auxiliary circuit breaker Q2 on and place the Tx in maintenance<br />
mode.<br />
5. Calibrate the +5VDC, +15VDC, -15VDC, +28VDC, and 40 VDC AND Ion pump voltage<br />
according to meter calibration procedures No.5 to No.9 as described in 6.3 of MEHB 10-<br />
511.<br />
72