Click the “Calculate” button to calculate the results. The simulation window will update<br />

with a rough graph, and the “Results” column at the right of the screen will have values for the<br />

cell and syringe concentrations. The calculated C value is listed below; its background is colorcoded.<br />

This value is equal to 5 in “Minimum Protein” mode, 100 in “Highest Quality” mode,<br />

and will vary if the user chooses to change cell concentration manually. The c-value predicts the<br />

shape or sigmoidicity of the curve. Optimal values for C are between 5 and 500 (green); values<br />

between 1 and 5, and 500 and 1000 should work but may not give the best result (yellow). C<br />

values less than 1 or greater than 1000 will probably not yield usable data (red). The user may<br />

adjust the two experimental concentrations by using the change buttons beside each<br />

concentration box. Once the cell concentration has been changed, clicking “Calculate” again<br />

will update the suggested syringe concentration relative to the new cell concentration.<br />

Any warnings will appear in the status bar near the top of the screen. These include<br />

“High heat warning”, meaning that the estimated heat that will be generated is more than the<br />

instrument can detect.<br />

It is highly recommended that the users look carefully at the projected curve and make<br />

sure that the shape and rough values are reasonable before proceeding.<br />

A pair of options at the bottom of this tab allows users to work in K d or K a , and to choose<br />

whether to view the simulation plot using raw heat per injection (DH) or the heat normalized to<br />

the molar ratio (NDH).<br />