3.2 Changing Injection Syringes<br />
Clean and dry the syringe before removal. Remove sample tube form loading hole and<br />
disconnect the cleaning apparatus. Be very careful whenever the syringe is not secured in the<br />
pipette, as the needle bends extremely easily. See images 1-7 below for illustrated removal.<br />
Slide the pipette to the cleaning position and insert it firmly into the hole. Turn the inner<br />
metal syringe holder to the right several full rotations. This will unscrew the metal ring that<br />
holds the syringe in the pipette. One the ring is loose, move the pipette to the empty loading<br />
hole. Use the manual pipette controls to lower the pipette tip about 0.3 inches. Without the<br />
bottom ring, the syringe will move down with the pipette tip.<br />
1.<br />
2. 3. 4.<br />
5. 6. 7. 8.<br />
9. 10. 11. 12.<br />