3.2 Changing Injection Syringes<br />

Clean and dry the syringe before removal. Remove sample tube form loading hole and<br />

disconnect the cleaning apparatus. Be very careful whenever the syringe is not secured in the<br />

pipette, as the needle bends extremely easily. See images 1-7 below for illustrated removal.<br />

Slide the pipette to the cleaning position and insert it firmly into the hole. Turn the inner<br />

metal syringe holder to the right several full rotations. This will unscrew the metal ring that<br />

holds the syringe in the pipette. One the ring is loose, move the pipette to the empty loading<br />

hole. Use the manual pipette controls to lower the pipette tip about 0.3 inches. Without the<br />

bottom ring, the syringe will move down with the pipette tip.<br />