3.6 Temperature Calibration Check<br />

It is recommended that the ITC temperature calibration be checked approximately once<br />

every 12 months. The measured temperatures should be within +/- 0.2 degrees of the set<br />

temperatures in the iTC 200 software. This procedure requires a Control Company digital meter<br />

and submersible probe or equivalent. The Control Company meter is model #9612; the probe is<br />

model #4021.<br />

Make sure the cells are clean, and load them with degassed, distilled water. Residue in<br />

the cells or empty cells will yield unpredictable results. Insert the probe into the sample cell and<br />

turn it on.<br />

1. Use the “Temperature” control box under the “Instrument Controls” tab in the iTC 200<br />

software to set the temperature to 30°C.<br />

2. Set the Plot Idle Data flag so that the current data is displayed.<br />

3. When the instrument is thermostatting at the target temperature, the DP has settled at a<br />

negative, stable value, and the DT is 0, use the data reader tool and the display from the<br />

meter to take temperature data points, as close together as possible.<br />

4. Repeat for the second check point at 70°C. If the error for each is not less than 0.2°C, please<br />

contact MicroCal for assistance.<br />

It is recommended that users keep a log with the date of the calibration check and the software<br />

and meter readings for each temperature point.<br />