5.1.6.3 Reference samples<br />

The substrate reference sample for non-conducting coating thickness<br />

measurement should have the same conductivity, surface contour and<br />

thickness as the material to be tested. For reference samples for the coating,<br />

stiff plastic shims of appropriate thicknesses are required. If the measurement<br />

is a go/no-go type, to verify whether the thickness lies within acceptable limits,<br />

only two shim thicknesses are necessary, one with the thickness of the lower<br />

limit and one with the thickness of the higher limit. However, a third shim, with<br />

thickness in the middle of the range, is desirable. If, however, the actual<br />

thickness of the coating is to be measured, at least three shim thicknesses<br />

are required. These should have thicknesses equal to the maximum and<br />

minimum expected thicknesses, and one approximately midway between<br />

these. If the coating required to be measured is thin, bare metal may be used<br />

for calibration. More shim thicknesses may be required if the response is<br />

notably nonlinear.<br />