5.1.7.5 Measurement procedure<br />

(a) Connect the probe to be used to the instrument, switch on the test<br />

instrument and allow it to warm up for at least 5 minutes.<br />

(b) Ensure the test area is clean and free from dirt or other contaminants.<br />

(c) Select the frequency required, and set the gain to the middle of its range.<br />

(d) Apply the probe to the thickest reference sample, balance the<br />

instrument, and adjust the phase so that the lift-off signal is horizontal to<br />

the left.<br />

(e) With the probe still on the sample, adjust the spot position to a location<br />

towards the lower right hand corner of the display (see FIG. 5.27).<br />

(f) Apply the probe to the thinnest reference sample and check whether a<br />

signal is displayed. Do not rebalance. If visible, the spot should lie<br />

above the previously obtained signal. Adjust the gain so that the spot is<br />

located towards the upper right hand corner of the screen. If the spot is<br />

not visible, first reduce the gain until it becomes visible, then further<br />

adjust the gain to locate the spot towards the upper right hand corner of<br />

the screen, as shown in FIG. 5.27.<br />

Charlie Chong/ Fion Zhang