21.12.2014 Views

Electromagnetic Testing - Eddy Current Testing Applications Chapter 5 & 6

Eddy Current Chapter 5 and 6

Eddy Current Chapter 5 and 6

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

5.1.3.4 Test procedure<br />

The general procedure for detection of surface flaws is as follows<br />

Standardisation:<br />

(a)<br />

(b)<br />

(c)<br />

(d)<br />

(e)<br />

Connect the probe to be used to the instrument, switch on the test<br />

instrument, conduct a battery condition check and allow the instrument to<br />

warm up for at least 5 minutes.<br />

Ensure the test area is clean and free from loose paint, dirt, or other<br />

contaminants.<br />

Select the frequency required and set the gain to the middle of its range.<br />

Apply the probe to the reference block at a location away from the<br />

reference discontinuities with the coil axis at right angles to the surface,<br />

and balance the instrument. Adjust the spot if required to approximately<br />

the centre of the display. The spot can be located in the lower right hand<br />

quadrant if desired to allow more space to view signals to the right (lift-off)<br />

and upwards (flaw signals).<br />

Raise the probe slightly from the surface to obtain a lift-off signal, and<br />

adjust the phase so that increasing lift-off gives a signal horizontally to<br />

the left, continually raising and lowering the probe to the test surface to<br />

monitor the lift-off signal during this adjustment.<br />

Charlie Chong/ Fion Zhang

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!