<strong>TO</strong> <strong>00</strong>-<strong>20</strong>-<strong>14</strong> <strong>30</strong> September <strong>20</strong>118.<strong>14</strong>.1.5.2 A portable or central vacuum cleaning system is required to clean TMDE and for general housecleaning.Cleaning TMDE within the calibration/repair area is acceptable, but is restricted to use of vacuum, small brushes, andsmall hand-held nontoxic spray-and-wipe cleaning. Where toxic fumes, excessive dust, or other safety hazards exist,cleaning shall be accomplished in an approved area outside the calibration and repair area. This includes devicescontaining mercury, such as manometers and barometers.NOTEEating or drinking in the Type IV is permitted only in a designated area set asidefor that purpose. A designated eating area is not authorized in the Type IV if anexisting break area is convenient. No calibration or repair is permitted in thatarea.8.<strong>14</strong>.2 F-15 Supporting Type II () PMEL Requirements.8.<strong>14</strong>.2.1 A 16' X 15' floor area is required within the calibration/ repair area to accommodate a 6' X 3' granite surfaceplate and a line-of-site area for optical calibration of a HUD Mount Alignment Adapter. Floor loading beneath thesurface plate shall be capable of supporting the surface plate (3,<strong>00</strong>0 pounds), a granite angle block (1,5<strong>00</strong> pounds) plusother ancillary equipment. Total loading is for 5,<strong>00</strong>0 pounds. The surface plate is mounted on either a tripod or aquadruped stand with four inch rectangular feet. Casters are normally provided on the stand.8.<strong>14</strong>.2.2 An increase in workload of approximately 3<strong>00</strong> items of F-15 related support equipment could be expected in theType II.8.<strong>14</strong>.3 F-16 Type IV PMEL Requirements.8.<strong>14</strong>.3.1 Electrical Power.a. 2<strong>00</strong> wye/115 +3-7 VAC, 3 Phase, 4 wire, 4<strong>00</strong> ±<strong>20</strong> Hz, 10 KVAb. 1<strong>20</strong> VAC ±10% 1 Phase, 3 wire, 60 ±6 or 50 ±3 Hz, <strong>20</strong> KVA.c. 240 VAC ±10%, 1 Phase, 3 wire, 60 ±6 or 50 ±3 Hz, 10 KVA.d. Grounding shall be in accordance with AFMAN 32-1094, measurement techniques per <strong>TO</strong> 31-10-24.8.<strong>14</strong>.3.2 Temperature and Humidity Controls. Temperature shall be maintained between 73 ±9 °F and humidity shallbe maintained between 15 and 70% RH. These temperature and humidity limits differ from those of the Type II PMEL inorder to permit use of the least restrictive facility requirements. These limits allow full use of the equipment.Temperature and humidity shall be continuously monitored and recorded.8.<strong>14</strong>.3.3 Floor Space. Minimum area for one set is 1040 ft 2 , two sets is 1,6<strong>00</strong> ft 2. . If the Type IV is consolidated intothe Type II, this is the additional space that could be required in the Type II.8.<strong>14</strong>.3.4 Lighting. (See paragraph 8.8.)8.<strong>14</strong>.3.5 Additional Requirements.8.<strong>14</strong>.3.5.1 Capping of unmated connectors is not required.8.<strong>14</strong>.3.5.2 A portable or central vacuum cleaning system is required to clean TMDE and for general housecleaning.Cleaning TMDE within the calibration/repair area is acceptable, but is restricted to use of vacuum, small brushes, andsmall hand-held nontoxic spray-and-wipe cleaning. Where toxic fumes, excessive dust, or other safety hazards exist,cleaning shall be accomplished in an approved area outside the calibration repair area. This includes devices containingmercury, such as manometers and barometers.NOTEEating or drinking in the Type IV is permitted only in a designated area set asidefor that purpose. A designated eating area is not authorized in the Type IV if anexisting break area is convenient. No calibration or repair is permitted in that area.8-8
<strong>30</strong> September <strong>20</strong>11 <strong>TO</strong> <strong>00</strong>-<strong>20</strong>-<strong>14</strong>8.<strong>14</strong>.4 F-16 Supporting Type II( ) PMEL Requirements.8.<strong>14</strong>.4.1 Type II( ) PMELs supporting the F-16 may require additional floor space to accommodate a 3’X6'surface plateand stand. A surface plate of this size shall require a working area approximately 15'X16' feet. Floor loading beneathsurface plate shall be capable of supporting approx. 5<strong>00</strong>0 pounds (35<strong>00</strong> pound surface plate plus ancillary blocks).8.<strong>14</strong>.4.2 Additional workspace may be required in a Type II PMEL supporting an F-16 ESS. An increase in workload ofapproximately 2<strong>00</strong> items of F-16 related support equipment can be expected in the Type II.8.15 ENVIRONMENTAL LIMITS FOR USNS OBSERVATION ISLAND AND RASCAL.8.15.1 USNS Observation Island environmental operating limits are as follows:a. Temperature = 73 ±6 °F.b. Humidity = 15% RH to 70% RH.8.15.2 RASCAL environmental operating limits are as follows:a. Temperature = 73 ±6 °F.b. Humidity = <strong>20</strong>% to 50% RH.8-9/(8-10 Blank)