Optimizing PerformanceUsing the Dual Power Meter DisplayFigure 5-6 Running Calibration(s) Bar (Zeroing Sensor)For details oneach key, use keyhelp as describedon page 42.NOTEThe U2000 Series USB Power Sensor, does not require a 50 MHz calibration. If acalibration is attempted with the U2000 Series Power Sensors, the <strong>MXG</strong> displays amessage reading:The U2000 series power sensor does not require a 50 MHz calibration. Refer toFigure 5- 7 on page 86.Figure 5-7 Diagnostic Dialog Box Displayed for U2000 Power SensorFor details oneach key, use keyhelp as describedon page 42.5. Press Return > Return > Channel A to OnThe current power meter sensor reading should be displayed in the ChA portion of theinstrument's display and in the upper right portion of the display under Power Meter. Refer toFigure 5- 8.86 Agilent <strong>N5161A</strong>/<strong>62A</strong>/<strong>81A</strong>/<strong>82A</strong>/<strong>83A</strong> <strong>MXG</strong> <strong>Signal</strong> <strong>Generators</strong> User’s <strong>Guide</strong>