Optimizing PerformanceUsing the Dual Power Meter Display13. On the <strong>MXG</strong>: Press Channel B to On and then back to Off again. This initializes the <strong>MXG</strong> to theexternal power meter.14. Press Return > Zero SensorA diagnostic dialog box is displayed each time an external power meter is being used and theZero Sensor or Calibrate Sensor softkey is pressed (refer to Figure 5- 10 on page 88).Verify the power sensor is connected to the 50 MHz reference of the power meter.Figure 5-10 Diagnostic Dialog Box for Channel B15. Press DoneThe Running Calibration(s) bar is displayed: Zeroing Sensor Please wait....16. After Running Calibration(s) bar disappears: Press Calibrate SensorDiagnostic Dialog box is displayed that prompts for verifying the connection of the power sensorto the power meter 50 MHz reference (refer to Figure 5- 11 on page 88).Figure 5-11 Diagnostic Dialog Box for CalibrationFor details oneach key, use keyhelp as describedon page 42.For details oneach key, use keyhelp as describedon page 42.88 Agilent <strong>N5161A</strong>/<strong>62A</strong>/<strong>81A</strong>/<strong>82A</strong>/<strong>83A</strong> <strong>MXG</strong> <strong>Signal</strong> <strong>Generators</strong> User’s <strong>Guide</strong>