Basic Digital Operation (Option 651/652/654)Modifying a FIR Filter Using the FIR Table EditorFigure 8-24For details on each key, use keyhelp as described on page 42.7. Press Return.Modifying the Coefficients1. Using the front panel arrow keys, highlight coefficient 15.2. Press 0 > Enter.3. Press Display Impulse Response.Figure 8-25For details on each key, use keyhelp as described on page 42.Refer to Figure 8- 25 on page 215. The graphic display can provide a useful troubleshooting tool(in this case, it indicates that a coefficient value is missing, resulting in an improper Gaussianresponse).4. Press Return.5. Highlight coefficient 15.6. Press 1 > Enter.Agilent <strong>N5161A</strong>/<strong>62A</strong>/<strong>81A</strong>/<strong>82A</strong>/<strong>83A</strong> <strong>MXG</strong> <strong>Signal</strong> <strong>Generators</strong> User’s <strong>Guide</strong> 215