CONELOG ® SCREW-LINE IMPLANT<br />

<strong>SURGICAL</strong><br />

PROCEDURE<br />

IMPLANTATION<br />

IMPLANT INSERTION<br />

The CONELOG ® SCREW-LINE implant is taken by the sterile handle and<br />

removed from the primary package (blister). Contamination from nonsterile<br />

parts must be avoided.<br />

46<br />

CAUTION!<br />

Before inserting the implant, the silicone plug and the CONELOG ® cover<br />

screw must be removed from the handle.<br />

Taking the implant by the handle, insert it into the implant bed. Make sure<br />

to observe the axial orientation of the implant bed. If a hole has been pretapped,<br />

the positions of the thread dog point in the cortical bone must<br />

match that on the implant.<br />

TIP: Carefully screw the implant to the left until you can feel the thread dog<br />

point. Then, using the handle, screw the implant to the right until it obtains<br />

sufficient grip to allow withdrawal of the handle.<br />

OPTIONAL ACCESSORIES:<br />

PICKUP INSTRUMENT<br />

If the intraoral space is insufficient to allow insertion of the implant with<br />

the handle, the implant can be removed from the handle using the PickUp<br />

instrument. Slide the PickUp instrument between the implant and handle<br />

onto the CONELOG ® insertion post and lift off the handle. for the insertion<br />

procedure, place the selected driver on the CONELOG ® insertion post. The<br />

implant is inserted into the bone, and the PickUp instrument is removed.<br />

Inserting the implant