CONELOG ® SCREW-LINE IMPLANT<br />
<strong>SURGICAL</strong><br />
PROCEDURE<br />
IMPLANTATION<br />
IMPLANT INSERTION<br />
The CONELOG ® SCREW-LINE implant is taken by the sterile handle and<br />
removed from the primary package (blister). Contamination from nonsterile<br />
parts must be avoided.<br />
46<br />
CAUTION!<br />
Before inserting the implant, the silicone plug and the CONELOG ® cover<br />
screw must be removed from the handle.<br />
Taking the implant by the handle, insert it into the implant bed. Make sure<br />
to observe the axial orientation of the implant bed. If a hole has been pretapped,<br />
the positions of the thread dog point in the cortical bone must<br />
match that on the implant.<br />
TIP: Carefully screw the implant to the left until you can feel the thread dog<br />
point. Then, using the handle, screw the implant to the right until it obtains<br />
sufficient grip to allow withdrawal of the handle.<br />
OPTIONAL ACCESSORIES:<br />
PICKUP INSTRUMENT<br />
If the intraoral space is insufficient to allow insertion of the implant with<br />
the handle, the implant can be removed from the handle using the PickUp<br />
instrument. Slide the PickUp instrument between the implant and handle<br />
onto the CONELOG ® insertion post and lift off the handle. for the insertion<br />
procedure, place the selected driver on the CONELOG ® insertion post. The<br />
implant is inserted into the bone, and the PickUp instrument is removed.<br />
Inserting the implant