ACCESSORIES<br />

CONELOG ® ADAPTER FOR NARROW GAPS AND<br />

HOLDING SLEEVE FOR SCREW IMPLANTS TO GUIDE<br />

THE ADAPTER INTO THE IMPLANT<br />

Two special reduced-diameter CONELOG ® adapters (long) are available for<br />

CONELOG ® SCREW-LINE implant diameters Ø 3.3 mm and 3.8/4.3 mm to<br />

assist manual insertion into small gaps. Three lateral marks match the cams<br />

on the adapter, thus providing a buccal orientation for one of the CONELOG ®<br />

SCREW-LINE implant grooves. After removing from the primary package<br />

(blister), the color-coded holding sleeve with the correct diameter is inserted<br />

over the endosseous part of the implant to guide the adapter into<br />

the screw implant. The sleeve is squeezed together like a collet at the level<br />

of the implant shoulder, and the handle is removed.<br />

NOTE<br />

We recommend using a hemostatic clip to squeeze the sleeve together<br />

to hold the implant securely and tightly in the sleeve. The handle can<br />

then be removed, the retention screw of the insertion post loosened<br />

with a screwdriver, hex, and the insertion post removed.<br />

IMPORTANT NOTE<br />

The CONELOG ® adapter and the holding sleeve must be sterilized before<br />

use.<br />

Setting up the<br />

sleeve<br />

Squeezing together sleeve with a<br />

hemostatic clip and removing handle<br />

Unscrewing the<br />

retention screw<br />

CONELOG ® SCREW-LINE IMPLANT<br />

CONELOG ® adapter, long,<br />

for Ø 3.3 mm and Ø 3.8/4.3 mm<br />

Holding sleeve for screw implants, Ø 3.3 mm,<br />

Ø 3.8 mm and Ø 4.3mm<br />

Removing the CONELOG ®<br />

insertion post<br />

51