CONELOG ® SCREW-LINE<br />

IMPLANTS<br />

INTRODUCTION<br />

CONELOG ® SCREW-LINE implants are endosseous implants available in<br />

various lengths and diameters. The implants are surgically inserted in the<br />

bone of the maxilla and/or mandible and serve as an anchor for functional<br />

and esthetic oral restorations for partially and fully edentulous patients. The<br />

prosthetic restoration is performed with single crowns, bridges or full dentures<br />

that are attached to the CONELOG ® implants with the appropriate<br />

CONELOG ® components. In principle, there are no preferable areas of insertion.<br />

The CONELOG ® SCREW-LINE implant is a geometrically tapered, self-tapping<br />

screw implant and is available with Promote ® surface.<br />

IMPLANTATION<br />

The CONELOG ® SCREW-LINE implant is not only suitable for late implantation,<br />

but for immediate or delayed immediate implantation as well. The<br />

healing technique can be submerged or transgingival. The implant is easily<br />

inserted because the taper of the implant body (3°–9° depending on length<br />

and diameter) induce self centering. The self-cutting thread provides for<br />

continuous grip on the bone and high primary stability.<br />

IMPORTANT NOTE<br />

The procedure for preparing the implant bed for CONELOG ® SCREW-LINE<br />

implants is identical to that performed with CAMLOG ® SCREW-LINE implants.<br />

The implant bed is prepared using SCREW-LINE drills and instruments.<br />

RANGE OF USE FOR CONELOG ® SCREW-LINE IMPLANTS:<br />

A more deeply positioned coronal implant shoulder is recommended particularly<br />

where a high-esthetic outcome is important. The CONELOG ®<br />

SCREW-LINE implant, Promote ® plus, has an acid-etched tapered implant<br />

shoulder (45°). The following clinical requirements should be met:<br />

• Normal to thick biotype<br />

• Gingival height of at least 3.0 mm<br />

• Minimum width of 1.0 mm of the attached gingiva<br />

• Minimum distance of 2.0 mm between attached gingiva and<br />

mimetic musculature<br />

CONELOG ® SCREW-LINE Implant<br />

Promote ® plus<br />

Machined implant<br />

shoulder surface<br />

CONELOG ® SCREW-LINE IMPLANT<br />

Acid-etched tapered implant<br />

shoulder (45°)<br />

Height: 0.1–0.2 mm (variable<br />

according to implant Ø)<br />

Abrasive-blasted,<br />

acid-edged Promote ®<br />

surface (Micro-Macro surface)<br />