All electro<strong>de</strong>s had very short adjustment times in 0,0001 M NH4 + even after 5 days of dry<br />

storage. Contrary to former measurements with Nitrate selective electro<strong>de</strong>s very little<br />

adjustment problems were found for all Ammonium (NH4 + ) selective electro<strong>de</strong>s even in the<br />

presence of interfering ions like Sodium (Na + ) or Potassium (K + ) at concentrations similar to<br />

those to be expected in seawater.<br />

RESULTS FOR TESTING THE NITRITE ION SELECTIVE<br />

ELECTRODES:<br />

The drifting behaviour observed during the adjustment time measurements also influences the<br />

measurements of the concentration series.<br />

Period of dry storage pH Electro<strong>de</strong> Adjustment time/ minutes<br />

706<br />

1 day 6.2<br />

4 days 6.0<br />

REAGECON 09 12<br />

REAGECON 10 12<br />

ORION 55<br />

REAGECON 09 14<br />

REAGECON 10 15<br />

ORION 50<br />

The interference of the chlori<strong>de</strong> ion is significant in the whole concentration range but much<br />

smaller than in the case of the nitrate and ammonium electro<strong>de</strong>s. The chlori<strong>de</strong> interference of<br />

the Reagecon electro<strong>de</strong>s is smaller than the influence on the Orion electro<strong>de</strong>.<br />

In the concentration range to be expected in seawater the interference of Bromi<strong>de</strong> (Br - ) and<br />

Nitrate (NO3 - ) is negligible at both temperatures. At 20 o C the interference of Sulphate (SO4 2- )<br />

is mo<strong>de</strong>rate, at 10 o C the influence of Sulphate (SO4 2- ) is small.<br />

At 10 o C interfering ions had less of an impact than at 20 o C.<br />

At pH 4.5 the slope of the Reagecon and the Orion electro<strong>de</strong> are satisfactory. The slopes of the<br />

REAGECON electro<strong>de</strong>s are somewhat closer to the theoretical value (Nernst value). At pH 8.5<br />

the curves differ significantly from the values measured at pH 4.5. Slope and linearity of the<br />

Orion electro<strong>de</strong> were very poor in the whole concentration range. In the range −<br />

NO2<br />

m < 0.001<br />

mol/kg the slope of the REAGECON electro<strong>de</strong>s differs significantly from the theoretical value<br />

(Nernst value).<br />

RESULTS FOR TESTING THE REFERENCE ELECTRODES<br />

CONSTRUCTION AND COMOPSITION:<br />

Reference electro<strong>de</strong>s of a large variety of construction and composition reflecting the current<br />

state of the art and proprietary knowledge have been prepared and challenged for speedy<br />

response, reproducibility and particularly long term stability in marine matrices and capable of<br />

withstanding bio-fouling. Depending on the outcome of subsequent specialised biological and<br />

physical testing there are 2 prototype reference electro<strong>de</strong> <strong>de</strong>signs which can be proceed with.